Dissolved Oxygen

Reference Designator
GS01SUMO-RID16-06-DOSTAD000
Start Depth
12
End Depth
12
Location
Near Surface Instrument Frame
Current Status
Unknown
Class
DOSTA (Dissolved Oxygen Stable Response)
Series
DOSTA-D
Science Discipline
Chemical
Make
Aanderaa
Model
Optode 4831
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/GS01SUMO/RID16/06-DOSTAD000/metadata
Method Data Stream Content Type
recovered_host dosta_abcdjm_dcl_instrument_recovered Data Products Report M2M Stats Science
telemetered dosta_abcdjm_dcl_instrument Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review AT-26-29 02/18/2015 12/27/2015 CGMGS-01SUMO-00001 CGINS-DOSTAD-00382 -54.4083 -89.3575 12 4611
2 Review NBP-15-11 12/14/2015 12/12/2016 CGMGS-01SUMO-00002 CGINS-DOSTAD-00227 -54.4041 -89.2069 12 4588
3 Review NBP16-10 11/25/2016 12/09/2018 CGMGS-01SUMO-00003 CGINS-DOSTAD-00428 -54.4076 -89.3567 12 4612
4 Review DY096 12/04/2018 CGMGS-01SUMO-00004 CGINS-DOSTAD-00507 -54.4072 -89.206 12 4589

No notes yet.

New Note

Metadata Start Date End Date Comment
GS01SUMO-RID16-06-DOSTAD000
7/25/15, 8:30 PM 8/24/15, 8:17 PM

Deployment 1: Dissolved oxygen noise increased, possibly from biofouling.

Id: 778 By: lgarzio
Flag: suspect Exclude: No

GS01SUMO-RID16-06-DOSTAD000
3/7/16, 7:30 PM 12/12/16, 2:58 AM

Deployment 2: Dissolved oxygen noise increased, possibly from biofouling.

Id: 779 By: lgarzio
Flag: suspect Exclude: No

GS01SUMO-RID16-06-DOSTAD000
11/24/16, 7:00 PM

The derived O2 products (L1 Temperature Compensated and L2 Salinity Corrected Dissolved Oxygen) are suspect, due to errors during vendor calibration. Until a correction can be applied, they should not be used from 11/25/2016 onwards.

Id: 85 By: lgarzio

GS01SUMO
12/8/17, 7:00 PM

Deployment 3: Recovery of the mooring was planned during the NBP 17-09 cruise (Nov - Dec 2017) but not completed due to adverse weather conditions. Recovery options are pending. Instruments that are still operational are likely reaching (or have surpassed) their deployment lifespan.

Id: 834 By: lgarzio