3-Wavelength Fluorometer

Reference Designator
CE07SHSM-RID27-02-FLORTD000
Start Depth
7
End Depth
7
Location
Near Surface Instrument Frame
Current Status
Unknown
Class
FLORT (3-Wavelength Fluorometer)
Series
FLORT-D
Science Discipline
Biological
Make
WET Labs
Model
ECO Triplet-w
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CE07SHSM/RID27/02-FLORTD000/metadata
Method Data Stream Content Type
recovered_host flort_sample Data Products Report M2M Stats Science
telemetered flort_sample Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review OC1503D 04/15/2015 09/23/2015 CGMCE-07SHSM-00001 CGINS-FLORTD-01152 46.9861 -124.566 7 87
2 Review OC1510A 10/13/2015 05/13/2016 CGMCE-07SHSM-00002 CGINS-FLORTD-01302 46.9875 -124.569 7 87
3 Review TN-342 05/05/2016 09/27/2016 CGMCE-07SHSM-00003 CGINS-FLORTD-01123 46.9895 -124.569 7 87
4 Review AT37-03 09/20/2016 04/13/2017 CGMCE-07SHSM-00004 CGINS-FLORTD-00996 46.9859 -124.566 7 87
5 Review SKQ201704S 04/11/2017 10/06/2017 CGMCE-07SHSM-00005 CGINS-FLORTD-00995 46.9882 -124.567 7 89
6 Review SKQ201715S 10/03/2017 03/29/2018 CGMCE-07SHSM-00006 CGINS-FLORTD-01291 46.9852 -124.564 7 87
7 Review SKQ201808S 03/25/2018 09/20/2018 CGMCE-07SHSM-00007 CGINS-FLORTD-01487 46.9881 -124.568 7 87
8 Review SR1813 09/18/2018 CGMCE-07SHSM-00008 CGINS-FLORTD-01154 46.9835 -124.564 7 87

No notes yet.

New Note

Metadata Start Date End Date Comment
CE07SHSM
12/1/17, 9:35 AM 3/25/18, 3:09 PM

Data loss due to a power system failure. The mooring stopped receiving power from photo voltaic panel 1 of 4 on Nov-2. By Nov-20, the mooring stopped receiving power from two more solar panels (PV2 & PV4) and one of two wind turbines (WT2). This indicates the Power System Controller (PSC) they connect to is failing. Mooring components will be shut down as the mooring loses the ability to recharge its batteries.

Id: 1237 By: crisien
Flag: not_operational Exclude: No