pCO2 Water

Reference Designator
CE07SHSM-MFD35-05-PCO2WB000
Start Depth
87
End Depth
87
Location
Multi-Function Node
Current Status
Unknown
Class
PCO2W (pCO2 Water)
Series
PCO2W-B
Science Discipline
Chemical
Make
Sunburst
Model
SAMI-pCO2
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CE07SHSM/MFD35/05-PCO2WB000/metadata
Method Data Stream Content Type
recovered_host pco2w_abc_dcl_instrument_blank_recovered Data Products Report M2M Stats Science
recovered_host pco2w_abc_dcl_instrument_recovered Data Products Report M2M Stats Science
recovered_host pco2w_abc_dcl_metadata_recovered Metadata Report M2M Engineering
recovered_host pco2w_abc_dcl_power_recovered Power Data Report M2M Engineering
recovered_inst pco2w_abc_instrument Data Products Report M2M Stats Science
recovered_inst pco2w_abc_instrument_blank Data Products Report M2M Stats Science
recovered_inst pco2w_abc_metadata Metadata Report M2M Engineering
recovered_inst pco2w_abc_power Power Data Report M2M Engineering
telemetered pco2w_abc_dcl_instrument Data Products Report M2M Stats Science
telemetered pco2w_abc_dcl_instrument_blank Data Products Report M2M Stats Science
telemetered pco2w_abc_dcl_metadata Metadata Report M2M Engineering
telemetered pco2w_abc_dcl_power Power Data Report M2M Engineering
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review OC1503D 04/15/2015 09/23/2015 CGMCE-07SHSM-00001 CGINS-PCO2WB-C0083 46.9861 -124.566 87 87
2 Review OC1510A 10/13/2015 05/13/2016 CGMCE-07SHSM-00002 CGINS-PCO2WB-C0119 46.9875 -124.569 87 87
3 Review TN-342 05/05/2016 09/27/2016 CGMCE-07SHSM-00003 CGINS-PCO2WB-C0147 46.9895 -124.569 87 87
4 Review AT37-03 09/20/2016 04/13/2017 CGMCE-07SHSM-00004 CGINS-PCO2WB-C0082 46.9859 -124.566 87 87
5 Review SKQ201704S 04/11/2017 10/06/2017 CGMCE-07SHSM-00005 CGINS-PCO2WB-C0083 46.9882 -124.567 89 89
6 Review SKQ201715S 10/03/2017 03/29/2018 CGMCE-07SHSM-00006 CGINS-PCO2WB-C0085 46.9852 -124.564 87 87
7 Review SKQ201808S 03/25/2018 09/20/2018 CGMCE-07SHSM-00007 CGINS-PCO2WB-C0052 46.9881 -124.568 87 87
8 Review SR1813 09/18/2018 CGMCE-07SHSM-00008 CGINS-PCO2WB-C0063 46.9835 -124.564 87 87

No notes yet.

New Note

Metadata Start Date End Date Comment
CE07SHSM-MFD35-05-PCO2WB000
Method: telemetered
Stream: pco2w_abc_dcl_instrument
4/15/15, 1:15 PM 1/9/19, 7:00 PM

UPDATED, 2019-01-15: After careful review, the original data product algorithm (pco2_pco2wat) for this sensor was found to contain an error that resulted in calculation of incorrect derived L1 pCO2 values. In consultation with the vendor, the algorithm has been corrected, verified, and uploaded to the OOI system. This update, effective 2019-01-10, applies to all deployments for this sensor. All derived L1 pCO2 data products generated prior to this date should be re-requested in order to ensure utilization of the correct algorithm. The updated algorithm can be viewed at https://github.com/oceanobservatories/ion-functions/blob/master/ion_functions/data/co2_functions.py. * Until further notice, derived L1 pCO2 data products from this sensor are incorrect and should not be used. A significant discrepancy has been uncovered between the pCO2 values from shipboard water samples vs. values calculated by OOI system, and the algorithms are under review: https://github.com/ooici/ion-functions/blob/master/ion_functions/data/co2_functions.py. To receive instant notification of updates to the code, add yourself as a watcher to the GitHub repository linked above. For further questions please email help@oceanobservatories.org.

Id: 277 By: friedrich.knuth

CE07SHSM-MFD35-05-PCO2WB000
Method: recovered_host
Stream: pco2w_abc_dcl_instrument_recovered
5/5/16, 3:35 PM 1/9/19, 7:00 PM

UPDATED, 2019-01-15: After careful review, the original data product algorithm (pco2_pco2wat) for this sensor was found to contain an error that resulted in calculation of incorrect derived L1 pCO2 values. In consultation with the vendor, the algorithm has been corrected, verified, and uploaded to the OOI system. This update, effective 2019-01-10, applies to all deployments for this sensor. All derived L1 pCO2 data products generated prior to this date should be re-requested in order to ensure utilization of the correct algorithm. The updated algorithm can be viewed at https://github.com/oceanobservatories/ion-functions/blob/master/ion_functions/data/co2_functions.py. * Until further notice, derived L1 pCO2 data products from this sensor are incorrect and should not be used. A significant discrepancy has been uncovered between the pCO2 values from shipboard water samples vs. values calculated by OOI system, and the algorithms are under review: https://github.com/ooici/ion-functions/blob/master/ion_functions/data/co2_functions.py. To receive instant notification of updates to the code, add yourself as a watcher to the GitHub repository linked above. For further questions please email help@oceanobservatories.org.

Id: 303 By: friedrich.knuth

CE07SHSM-MFD35
8/11/16, 8:00 PM 9/26/16, 8:00 PM

The Power System Controller on CE07SHSM indicates the power connection to the MFN short-circuited on 2016-08-12 at 23:30 UTC. Failure is indicated by decreased voltage (70 VDC instead of 380 VDC) and increased current draw (1900 mA instead of 120 mA) to the CVT. This symptom is indicative of a short-circuit in the power lines of either the stretch hose (most likely) or EM Chain. In the past, this failure has been associated with a failed stretch hose. Have disabled all further power and communications with the MFN. This means no more telemetered data from: ADCPT2, VEL3D, CTDBP2 + DOSTA, PRESF, PCO2W, PHSEN2, OPTAA2, ZPLSC

Id: 908 By: michaesm

CE07SHSM-MFD35-05-PCO2WB000
8/11/16, 8:00 PM 9/26/16, 8:00 PM

The Power System Controller on CE07SHSM indicates the power connection to the MFN short-circuited on 2016-08-12 at 23:30 UTC. Failure is indicated by decreased voltage (70 VDC instead of 380 VDC) and increased current draw (1900 mA instead of 120 mA) to the CVT. This symptom is indicative of a short-circuit in the power lines of either the stretch hose (most likely) or EM Chain. In the past, this failure has been associated with a failed stretch hose. Have disabled all further power and communications with the MFN. Therefore, there is no telemetered data for this reference designator.

Id: 913 By: michaesm
Flag: not_available Exclude: No

CE07SHSM-MFD35-05-PCO2WB000
10/12/16, 8:00 PM 4/12/17, 8:00 PM

DCL35 stopped reporting data on 2016-10-13 04:26 UTC. The MPEA indicates the system is booting up and drawing current, but it is not fully booting up and powering on the instruments. There will no longer be any telemetered data for this instrument.

Id: 914 By: michaesm
Flag: not_available Exclude: No

CE07SHSM
12/1/17, 9:35 AM 3/25/18, 3:09 PM

Data loss due to a power system failure. The mooring stopped receiving power from photo voltaic panel 1 of 4 on Nov-2. By Nov-20, the mooring stopped receiving power from two more solar panels (PV2 & PV4) and one of two wind turbines (WT2). This indicates the Power System Controller (PSC) they connect to is failing. Mooring components will be shut down as the mooring loses the ability to recharge its batteries.

Id: 1237 By: crisien
Flag: not_operational Exclude: No