pCO2 Air-Sea

Reference Designator
CE02SHSM-SBD12-04-PCO2AA000
Start Depth
0
End Depth
0
Location
Buoy Well (1 probe in air, 1 probe in water)
Current Status
Unknown
Class
PCO2A (pCO2 Air-Sea)
Series
PCO2A-A
Science Discipline
Chemical
Make
Pro-Oceanus
Model
pCO2-pro
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CE02SHSM/SBD12/04-PCO2AA000/metadata
Method Data Stream Content Type
recovered_host pco2a_a_dcl_instrument_air_recovered In-Air Data Products Report M2M Stats Science
recovered_host pco2a_a_dcl_instrument_water_recovered In-Water Data Products Report M2M Stats Science
telemetered pco2a_a_dcl_instrument_air In-Air Data Products Report M2M Stats Science
telemetered pco2a_a_dcl_instrument_water In-Water Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review OC1503D 04/02/2015 09/22/2015 CGMCE-02SHSM-00001 CGINS-PCO2AA-33198 44.6383 -124.304 0.7 80
2 Review OC1510A 10/07/2015 05/16/2016 CGMCE-02SHSM-00002 CGINS-PCO2AA-35258 44.6354 -124.303 0.7 82
3 Review TN-342 05/17/2016 09/28/2016 CGMCE-02SHSM-00003 CGINS-PCO2AA-34196 44.6357 -124.304 0.7 80
4 Review AT37-03 09/26/2016 04/22/2017 CGMCE-02SHSM-00004 CGINS-PCO2AA-35257 44.6393 -124.304 0.7 80
5 Review SKQ201704S 04/20/2017 10/15/2017 CGMCE-02SHSM-00005 CGINS-PCO2AA-35259 44.6358 -124.304 0.7 81
6 Review SKQ201715S 10/10/2017 04/04/2018 CGMCE-02SHSM-00006 CGINS-PCO2AA-34199 44.6393 -124.304 0.7 80
7 Review SKQ201808S 04/03/2018 09/25/2018 CGMCE-02SHSM-00007 CGINS-PCO2AA-34196 44.6357 -124.304 0.7 80
8 Review SR1813 09/22/2018 CGMCE-02SHSM-00008 CGINS-PCO2AA-34199 44.6395 -124.301 0.7 80

No notes yet.

New Note

Metadata Start Date End Date Comment
CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
4/4/18, 2:58 PM 9/24/18, 11:56 AM

PCO2A air measurement for the Spring 2018 deployment is considered suspect. Recent investigations have indicated a vendor error in assembly may have pinched closed an internal tube carrying the surface air stream into the sample volume resultant in an erroneous measurement.

Id: 1446 By: wingardc
Flag: suspect Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
11/11/18, 3:00 AM

Suspect CO2 sensor internal to the instrument has failed. Sensor started reporting erroneous values for the CO2 measurements and then flat-lined to 0.

Id: 1538 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_water
11/11/18, 3:00 AM

Suspect CO2 sensor internal to the instrument has failed. Sensor started reporting erroneous values for the CO2 measurements and then flat-lined to 0.

Id: 1539 By: wingardc
Flag: fail Exclude: No